In te application of: Brauckman, Richard A. 

5 Serial No.: 09/858,366 Group Art Unit: 3732 

Filed: May 16, 2001 Examiner: Ramana, Anuradha 

For: CATHETER ATTACHMENT AND CATHETER FOR 
10 BRACHYTHERAPY 



15 DECLARATION OF DR. JOHN LOBDELL PURSUANT TO 37 C.F.R. $1.132 

Assistant Commissioner for Patents 
P.O. Box 1450 
Alexandria, VA 22313-1450 

20 

Sir: 

1. I, Dr. John Lobdell, hereby declare as follows: 

25 2. I am an employee of the Theragenics Corporation and have 3 6 years experience in the 

nuclear industry. My detailed curriculum vitae is attached hereto as Exhibit A. 

3. I have reviewed the specification, drawings and currently pending claims of U.S. patent 
application no. 09/858,366. I have also reviewed the Office Action mailed on May 20, 

30 2004 (hereinafter "the Office Action"). 

4. I am informed that claims 1 -6 of the present application stand rejected under 35 U.S.C. 
§ 1 1 2, l ?t paragraph, as failing to comply with the enablement requirement on the basis 
that, "It is the Examiner's position that 'sufficient bond strength' cannot be determined 

35 without undue experimentation." 

5. I am informed that the standard for determining whether the specification meets the 
enablement requirement was set forth itiMineral Separation v. Hyde, 242 U.S. 261, 270 
(1916), which posed the question, "is the experimentation needed to practice the 

4 q invention undue or unreasonable?" 



SEP 02 '04 13=50 FR THERRGENICS 



180045S4303 TO 912155990601 P. 03/07 



U.S. Patent application no.: 09/858,366 2 

6. I am also informed that the factors to be considered when determining whether there is 
sufficient evidence to support a determination that a disclosure does not satisfy the 
enablement requirement and whether any necessary experimentation is "undue" 

5 include, but are not limited to: 

(A) The breadth of the claims; 

(B) The nature of the invention; 
10 (C) The state of the prior art; 

(D) The level of one of ordinary skill; 

(E) The level of predictability in the art; 

15 

(F) The amount of direction provided by the inventor; 

(G) The existence of working examples; and 

20 (H) The quantity of experimentation needed to make or use the invention based 

on the content of the disclosure. 

In re Wands, 858 F.2d 731,737,8 USPQ2d 1400, 1404 (Fed, Cir. 1988) andMPEP 
§2164.01(a). 

25 

Claim 1 

7. With regard to the breadth of the claims at issue, claim 1 of the present application is 
relatively narrow in scope since it requires a radioactive source bonded to a surface of 
the distal section of an elongate, flexible catheter body with sufficient bond strength 

30 that under normal conditions of use of the catheter, this radioactive source will not 

detach from the catheter body when in use. 

8. With regard to the nature of the. invention, the invention relates to a catheter useful for 
radiation treatment, wherein the catheter has a radioactive source bonded to a surface of 

35 the distal section of the catheter such that when the catheter is used the bonded 

radioactive source does not detach from the catheter. Since catheters provided with 
radioactive material have been known for quite some time, as evidenced by U.S. Patent 
nos. 5,199,939 and.5,282,781, the nature of the invention is to provide an improvement 



SEP 02 '04 13:50 FR THERRGEN I CS 



18004584303 TO 912155990601 P. 04/07 



U.S. Parent application no.: 09/858,366 3 

to existing catheters with radioactive material. As such, the invention lies in the 
improved positioning and manner whereby the radioactive material is associated with 
the catheter. Accordingly, since the nature of the invention is not unduly complex and 
relates to the provision of an improvement of conventional devices, it is my opinion that 
the skilled person will have a high degree of knowledge regarding the subject matter of 
the invention. 

9. The state of the prior art is evidenced by, for example, two of the documents of record 
in the present application, namely, "U.S. Patent No. 5,199,939 to Dake et al. 
(hereinafter "the Dake ct al. patent"); and "U.S. Patent No. 5,302,1 68 to Hess 
(hereinafter "the Hess patent"). The state of the prior art is such that devices similar to 
the device of the present invention are known from the prior art. 

10. The level of ordinary skill in the art is high. Specifically, the user of catheters is, at 
minimal, highly trained medical personnel such as a medical doctor (MD), a doctor of 
veterinary medicine (DVM), or medical personnel under the direct supervision of such 
persons, or possibly a Ph.D. researcher. Often, the designers of catheters are the 
medical personnel themselves as they are most intimately acquainted with the 
advantages and disadvantages each type catheter has. Additionally, biomedical 
engineers who have received specialized training in the art may design medical devices. 

11. A very significant amount of direction is provided in the present application for 
implementation of the device of claim 1. For example, the specification explains that 
the radioactive source may be chemically or thermally bonded to the catheter. See page 
5 lines 16-17. The application also explains that suitable catheter materials are well 
known in the art, See page 6 lines 1-2. The application explains that suitable methods 
for bonding the radioactive source to the catheter wall include at least electroless 
plating, chemical vapor deposition, electroplating, ion implantation, sputtering, physical 
vapor deposition, and the application of a coating of a polymer matrix having material 
dispersed therein. See page 7, lines 20-24. 



SEP 02 '04 13=50 FR THERRGEN I CS 



18004584303 TO 912155990601 P. 05/07 



U.S. Patent application no,: 09/858,366 ' . 4 

12. In addition, beginning on page 14, line 32, the specification describes the use of a 
diluent for purposes of promoting strong adhesion of the radioactive source material to 
the substrate, thereby forming a physiological inert layer which will not allow the 
radioactive source to be mobilized into the circulation of the patient being treated. See 
page 1 5, lines 20-24, and page 1 6, lines 1-4. A description of the suitable substrates to 
which the radioactive source is bonded to, or incorporated into, begins on page 16, line 
10 and continues though page 17, line 18. The specification explains bonding the 
radioactive material to the substrate at page 17, line 19 to page 20, line 13. 

13. The level of predictability in the art of radioactive catheter manufacturing is high. 
From the teachings of the present specification, using the well-known standard testing 
methods discussed below, it is my opinion that the skilled person can make and use 
catheters in accordance with the subject matter of the claims of the above-identified 
application. 

14. The International Organization for Standardization (ISO), a worldwide federation of 
national standards bodies, publishes standards regulating sealed radioactive sources. 
The American National Standards Institute (ANSI) is the primary source and official 
sales agent in the United States for ISO. ISO has published an International Standard 
for testing sealed radioactive sources, ISO 2919 (copy attached as Exhibit B hereto). 
ISO 291 9 (2 nd Ed 1999-01-15) sets forth the general requirements and classification for 
sealed radioactive sources. ISO 2919 sets forth detailed performance requirements in 
Section 6 for a variety of radioactive sources, including those used for medical 
brachytherapy. In ISO 291 9, Table 4 provides for sealed source classification 
(performance) requirements for typical usage (see page 7). These sources are tested 
using the criteria for brachytherapy sources. 

15. Moreover, in Section 7 of ISO 29 19, detailed tests for determining whether medical 
brachytherapy sources meet the performance requirements of Section 6 of ISO 2919 are 
set forth. These tests include the temperature, pressure, impact, vibration and puncture 
tests. Each of these tests has, as one purpose, to determine whether under simulated 



SEP 02 '04 13=50 FR THERAGENICS 



18004584303 TO 912155990601 P. 06/07 



U.S. Patent application no.: 09/858366 5 

conditions of use, the radioactive source is bonded to the substrate in a manner whereby 
the radioactive source will not detach from the device when in use. Thus, ISO 29 1 9 
demonstrates that there are well-known, standardized tests for brachytberapy sources 
for detenmining ^sufficient bond strength" under simulation conditions of use including 
temperature, pressure, impact, vibration and puncture resistance. 

1 6. ISO 29 19 also refers to the requirement that radioactive sources must be tested for 
leakage and refers to ISO 9978 for this purpose (See Section 3.4 of ISO 2919). ISO 
9978 (copy attached as Exhibit C hereto) provides leakage test methods for radioactive 
sources. The tests set forth in ISO 9978 include immersion tests, gaseous emanation 
tests, wipe tests, helium mass spectrometer leakage tests, bubble leakage tests, and 
water pressurization tests, each of which tests is designed to simulate conditions that 
may be encountered in use of radioactive brachytherapy devices, such as contact with 
bodily fluids. Thus, ISO 9978 demonstrates that there are well-known standardized 
tests for determining if a radioactive source will detach from a substrate by leakage 
under simulated conditions of use. Brachytherapy sources are tested using the criteria 
presented in this standard. 

1 7. In my opinion, the skilled person is capable of determining "sufficient bond strength 1 ' 
using one or more of the tests set forth in the well-known international standards ISO 
2929 and ISO 9978, discussed above. 



1 8. The U.S. Food and Drug Administration (hereinafter "the FDA") regulates medical 
devices including radioactive catheters (under the direction of the Center for Devices 
and Radiological Health). For example the Center for Devices and Radiological Health 
issued a document entitled, "Guidance for Submission of Premarket Notifications for 
Photon-Emitting Brachytherapy Sources/ 1 on August 2, 2000 (copy attached as Exhibit 
D hereto) (hereinafter "FDA Guidance"). In the FDA Guidance, Section IIL2.B, the 
applicable guidance and standards for photon-emitting brachytherapy sources are listed. 
Among the listed standards are the ANSI N43.6-1997 Classification of Sealed 
Radioactive Sources and the ISO 2919 Sealed Radioactive Sources - General 



SEP 02 '04 13:51 FR THERAGENICS 



18004584303 TO 912155990601 P. 07/07 



U.S. Patent application no.: 09/858,366 6 

Requirements and Classification. This demonstrates that the FDA recognizes TSO 2919 
as providing suitable test methods for evaluation of brachytherapy devices. 

5 19. In my opinion, essentially no experimentation is required to implement the present 

invention based on the disclosure of the application as filed, taken in combination with 
the common general knowledge of a skilled person. As discussed above, the skilled 
person already knows how to test the bond strength for radioactive sources under 
typical usage conditions as regulated by the FDA under the guidance of the Center for 
10 Devices and Radiological Health (CDRH) using the tests set forth in TSO 2199 and ISO 

9978. 



Dependent Claims 

20. Dependent claims 2-6 further specify the type of catheter body, where the radioactive 
15 source is bonded to in relation to the surface of the catheter, a retractable sheath for 

shielding the radioactive material and the nature of the radioactive material itself. In 
my opinion, the skilled person can implement these dependent claims with essentially 
no experimentation by using general knowledge from the art for the same reasons as 
specified above. 

20 

21. T hereby declare that all statements made herein of my own knowledge are true and that 
all statements made on information and belief are believed to be true, and further that 
the statements were made with the knowledge that willful false statements and the like 
made are punishable by fine or imprisonment, or both, under Section 1001 of Title 18 

25 of the United States Code and that such willful false statements may jeopardize the 

validity of the application or any patent issued thereon. 




** TQTAL PAGE . 07 ** 




John L. Lobdell, Ph.D., CHP 

6205 Deerwoods Trail 
Alpharetta, GA 30005 
(H) 678-947-0004 
(W) 770-831-4240 
E-mail: lobdellj@theragenics.com 



Education: 

B.S. in Physics with a minor in Mathematics, Spring Hill College, Mobile, Alabama, 1964. 

M.S.P.H. in Radiological Hygiene, University of North Carolina at Chapel Hill, 1968. 

Ph.D. in Health Physics, Georgia Institute of Technology, 1995. Research Topic: "Dose 
Rate and Spectral Photon Measurements Around a Large B WR Using a Tissue Equivalent 
Plastic Scintillator. 1 ' Advisor: Dr. N. E. Hertel. 

Five-week class in Boiling Water Reactor Technology at Browns Ferry Nuclear Plant, 
1969. 

"Occupational & Environmental Radiation Protection", Harvard School of Public Health, 
August 19-23, 1985. 

"Health Physics in Radiation Accidents", Oak Ridge Associated Universities, September 
8-12, 1986. 

"Workshop on Measurement Quality Assurance for Ionizing Radiation", National Institute 
of Standards and Technology (NIST), March 16-18, 1993 

"Media Center Appearances", C. S. Armstrong Associates, Inc., September 8, 1994 

"Radioactive Waste Packaging, Transportation, and Disposal Workshop", Chem-Nuclear 
Systems, July 21-24, 1997, and May 22-26, 2000. 



Professional Certification: 

Certified in Health Physics by the American Board of Health Physics, 1972. Recertified in 
1981, 1985, 1989, 1993, 1997, and 2001. 

Lead Auditor as defined by ANSI N45.2.23-1978, "Qualification of Quality Assurance 
Program Audit Personnel for Nuclear Power Plants". 



1 



HAZMAT trained and certified to ship radioactive material, July 1997 through May 2003. 



Health Physics Work Experience: 

July 1964 to August 1966: 

Employed by the Alabama and Virginia Departments of Health to operate the counting 
rooms to determine the radioactive content of environmental samples. 

June 1968 to September 1996, employed by the Tennessee Valley Authority, Muscle Shoals, 
AL 

From June 1968 to December 1979, 1 supervised the operation of the following programs: 
environmental radiological monitoring around TVA's nuclear power plants, health physics 
training, applied health physics services, film badge and TLD personnel monitoring services, 
whole body counting, and calibration of portable radiation survey instrumentation. 

From December 1979 to May 1980, 1 coordinated within TV A the modification of the 
radiological emergency plan for all of TVA's operating nuclear power plants in compliance with 
NUREG-0654 "Criteria for Preparation and Evaluation of Radiological Emergency Response 
Plans and Preparedness in Support of Nuclear Power Plants". 

From June 1980 to May 1982, 1 supervised a Quality Assurance/ALARA Staff. The Staff 
provided quality assurance services to a large health physics organization. 

From June 1982 to April 1985, 1 was the Staff Health Physicist in the office of the Chief, 
Health Physics Services. I provided health physics expertise to the Chief and all sections within 
the organization. During a ten-month period, I was assigned to the Browns Ferry Nuclear Plant 
as the health physicist on a recirculation pipe replacement project on unit 1 . 

From May 1985 to October 1986, 1 supervised a Dosimetry Section that coordinated and 
provided direction for the internal and external dosimetry programs in TVA. 

From November 1986 to September 1996, 1 supervised a section that repaired, maintained, 
modified, and calibrated portable radiation survey instrumentation. 

January 1997 to August 2003: employed by Novoste Corporation, Atlanta, GA. 

Novoste is a company that has developed a vascular brachytherapy system for the reduction 
of restenosis in coronary arteries. I was the Principal Radiation Physicist. When I initiated 



employment, the company had no radioactive materials license. The company acquired a 
research and development license and a manufacturing broad scope license through the State of 
Georgia. I was the Radiation Safety Officer for these licenses. 

During this period, I accomplished the following: 

- planned and developed the laboratory that receives, tests, calibrates, and ships 90 Sr sources 

in transfer devices throughout the world, 

- performed all the test method validations, installation qualifications, and process hazard 

analyses for the laboratory, 

- developed the techniques for the calibration of the source trains traceable to the National 

Institute of Standards and Technology, 

- set up and developed all the procedures for the operation of the laboratory, 

- calculated doses rates from the 90 Sr sources using Monte Carlo computer codes, 

- designed shielding for the transfer devices using the Monte Carlo codes, 

- provided input to FDA on radiation issues, 

- developed and gave training programs for company personnel. 

During the period, I traveled extensively throughout the US and Western Europe to interface 
with users, partners, and regulatory agencies. 

September 2003 to Present: employed by Theragenics Corporation, Buford, GA 

During this period, I am the Director of Health Physics and the Radiation Safety Officer. 
Theragenics manufactures radioactive seeds containing 125 I and 103 Pd for the treatment of prostate 
cancer. 

June 1991 to 1995: 

I served as a Technical Expert for the National Voluntary Laboratory Accreditation 
Program (NVLAP) for the Secondary Calibration for Ionizing Radiation Laboratory 
Accreditation Program. 

Teaching Experience: September 1989 to August 1991 : 

I taught four subjects at Shoals Community College: physics with calculus, two classes in 
physics without calculus, and health physics for radiographers. I taught a total of 14 quarters. 

Significant Papers and Publications: 

"Suitability of Glass-Encapsulated CaF2:Mn Thermoluminescent Dosimeters for Environmental 



Radiation Surveillance", presented at the National Health Physics Society Meeting in Miami, 
June 1973. 

" A TLD System for Personnel Monitoring", presented at the meeting of the Deep South and 
Alabama Chapters of the Health Physics Society, Gulf Shores, Alabama, August 1977. 

"Training for a Viable Nuclear Power Plant Radiological Emergency Plan", presented at the 
► Thirteenth Midyear Topical Symposium of the Health Physics Society, Honolulu, December 

1979. 

"Health Physics Planning for Recirculation Pipe Replacement at a BWR", presented at the annual 
meeting of the American Nuclear Society, New Orleans, June 1 984. 

"Calibration of DMC-90s in TVA", presented at the Merlin Gerin User's Group Meeting, Atlanta, 
April 1992. 

"A Tissue Equivalent Detector Photon Response Matrix", presented at the winter meeting of the 
American Nuclear Society, San Francisco, October 1995. 

"Dose Rate And Spectral Photon Measurements Around A Large BWR" presented at the Annual 
Health Physics Society Meeting, Seattle, July 1996. 

"Scanning Personnel For Internal Deposition Of Radioactive Material With Personnel 
Contamination Whole Body Friskers And Portal Monitors" presented at the Annual Health 
Physics Society Meeting, Seattle, July 1996. 

"Photon Spectra and Dose Measurements Using a Tissue-Equivalent Plastic Scintillator", 
Radiation Protection Dosimetry, Vol.72, No.2, pp.95- 103 (1997). 

"Gamma Ray Dose Rate Measurements at a Boiling Water Reactor", Radiation Protection 
Dosimetry, Vol.74, No.3, pp.163-171 (1997). 

"Radiochromic Film Dosimetry of a High Dose Rate Beta Source for Intravascular 
Brachytherapy", Medical Physics, 26 (1 1), November 1999. 

"Attenuation of Dose From a 90 Sr/Y Line Source by a Stainless-Steel Stent", Second Annual 
Radiotherapy Conference to Reduce Restenosis, La Jolla, CA, January 16-17, 1998. 

"Dosimetry and Radiation Aspects of the Use of 90 Sr- 90 Y for Intravascular Brachytherapy", 
Council on Ionizing Radiation Measurement and Standards, Gaithersburg, MD, October 30 - 
November 1,2000. 

"Quality Control for Novoste 90 Sr- 90 Y Intravascular Brachytherapy Sources", Council on 
Ionizing Radiation Measurements and Standards, Gaithersburg, MD, October 29-3 1 , 2001 . 



4 



"Dose Distribution Superimposed on Cine Images Acquired During IVBT", Cardiovascular 
Radiation Therapy , February 6-8, 2002. 

"Display of Dose Distribution of Cine/Fluoro Images Acquired During IVBT", AAPM Annual 
Meeting, Montreal, Canada, July 14-18, 2002. 

"Dosimetric Characterization of a New 90 Sr/ 90 Y Source with Balloon for Intravascular 
Brachytherapy", AAPM Annual Meeting, Montreal, Canada, July 14-18, 2002. 